HELP + SUPPORT

It is important to read the LEVL Instruction Manual before using your LEVL device. The manual will walk you through everything you need to know about your LEVL device including set up, pairing with the optional app and taking a breath measurement. READ MORE for IMPORTANT TIPS FOR GETTING STARTED.

IMPORTANT TIPS FOR SUCCESS

  • The LEVL device should be placed in a stable environment away from certain chemicals found in a clinical setting, like rubbing alcohol, cleaning products, hand sanitizer, sani-cloth wipes, air fresheners, essential oils and other strong fragrants.
  • Once the LEVL device is connected to power, it will start a 24-hour power up sequence to prepare the sensor for breath measurements. Calibration lights will start blinking at 12 hours, but please allow a full 24 hours for the sensor to stabilize before performing the calibration sequence.
  • Calibration takes about 10 minutes, then your device is ready for use.
  • When preparing for a breath measurement, please wait 60 minutes before taking a LEVL measurement if you have been exposed to the following.  These items may cause false high measurements and, in some cases, may cause the sensor to fail.

Alcohol, mouthwash, toothpaste

Breath mints, gum, cough drops

Tobacco and e-cigarettes

Lip balm or lipstick (wipe off before the test)

Non-sugar sweeteners (e.g., Sorbitol), water enhancers (e.g., Mio)

HOW-TO VIDEOS

HOW-TO CALIBRATE YOUR LEVL DEVICE

HOW-TO REPLACE THE SENSOR MODULE

TAKING A BREATH MEASUREMENT